Indenter
img25
Click here for sample calibration certificate

Calibration details & Traceability:

The above Rockwell indenter is performance verified against reference indenter using a standardising machine at FMCTL. The standardising machine is directy calibrated as per the requirements of ISO 6508-3:2015 and ASTM E18-16 annex A2. The standardising machine is calibrated using devices traceable to NPL-India, NPL- UK, IMGC, NIST or PTB either directly or through NABL, UKAS, NVALP or DAkkS accredited laboratories.

Calibration Method:

FMCTL/SOP/Rockwell based on ISO 6508-2 and IS 1586 (Part 2) standards.
Measurement Results
Test Point Hardness Level Deviation from Reference Indenter Admissible Deviation Expanded Uncertainty
Test Point 1 HRC 25 +0.36 HRC ±0.8 HRC ±0.73 HRC
Test Point 2 HRC 63 -0.21 HRC ±0.8 HRC ±0.34 HRC
Test Point 3 HR30N 64 +0.04 HR30N ±0.8 HR30N ±0.56 HR30N
Test Point 4 HR15N 91 +0.07 HR15N ±0.8 HR15N ±0.44 HR15N

Conformance Statement: This indenter fulfils the requirements of performance verification in accordance with the standards: ISO 6508-2 & IS 1586 (Part 2).